International Journal of Clinical Biochemistry and Research

Print ISSN: 2394-6369

Online ISSN: 2394-6377

CODEN : IJCBK6

International Journal of Clinical Biochemistry and Research (IJCBR) open access, peer-reviewed quarterly journal publishing since 2014 and is published under auspices of the Innovative Education and Scientific Research Foundation (IESRF), aim to uplift researchers, scholars, academicians, and professionals in all academic and scientific disciplines. IESRF is dedicated to the transfer of technology and research by publishing scientific journals, research content, providing professional’s membership, and conducting conferences, seminars, and award more...

Article type

Original Article


Article page

267- 271


Authors Details

Nithya K N, Meera S*


Article Metrics


View Article As

 


Downlaod Files

   






Article statistics

Viewed: 1152

PDF Downloaded: 916


Sigma metrics - A guide to quality control strategy in clinical Biochemistry laboratory


Original Article

Author Details : Nithya K N, Meera S*

Volume : 7, Issue : 2, Year : 2020

Article Page : 267-271

https://doi.org/10.18231/j.ijcbr.2020.058



Suggest article by email

Get Permission

Abstract

Introduction and Aims: Clinical laboratories are backbone of health care system & the responsibility lies
in the laboratory to produce reliable, reproducible & timely interpreted test results. Presently Westgard
quality control(QC) rules are the basis for quality control validation of laboratories. A new quality
assessment system known as Six sigma is considered as world class quality & is a uniquely defined scale
with which one can assess the performance of a laboratory. An analytical procedure should achieve a good
sigma levels for high reliability for the results. Many laboratories are not aware about the usefulness of six
sigma. Hence this study was aimed to assess and to plan the quality control strategy.
Materials and Methods: In this study, the internal quality control data of urea, creatinine, sodium,
potassium & chloride analysed for a period of 6 months & used to calculate bias, coefficient of variation
(CV)%, mean & standard deviation(SD) for each levels(L1,L2).
Results: Sigma value calculated shown Highest for Urea & Potassium (Level1 & Level 2) COBAS 6000
an integrated modular analyser. Sigma of 3 is the minimum acceptable level of quality. The frequency of
internal QC should be increased and corrective action should be taken for those parameters with sigma <3> Sigma metrics of each parameter helped us to identify the right QC strategy for that analyte using Westgard
sigma rule.
Conclusion: Hence six sigma is a more efficient way to control quality & is a self-assessment tool for
quality improvement.

Keywords: Quality control, Six sigma, Standard deviation (SD), Total allowable error.


How to cite : Nithya K N, Meera S, Sigma metrics - A guide to quality control strategy in clinical Biochemistry laboratory. Int J Clin Biochem Res 2020;7(2):267-271

This is an Open Access (OA) journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.