International Journal of Clinical Biochemistry and Research

Print ISSN: 2394-6369

Online ISSN: 2394-6377

CODEN : IJCBK6

International Journal of Clinical Biochemistry and Research (IJCBR) open access, peer-reviewed quarterly journal publishing since 2014 and is published under auspices of the Innovative Education and Scientific Research Foundation (IESRF), aim to uplift researchers, scholars, academicians, and professionals in all academic and scientific disciplines. IESRF is dedicated to the transfer of technology and research by publishing scientific journals, research content, providing professional’s membership, and conducting conferences, seminars, and award more...

Article type

Null


Article page

28- 30


Authors Details

Ketan K Mangukiya, Mittal Panchal


View Article As

 


Downlaod Files

   






Article statistics

Viewed: 421

PDF Downloaded: 278


Impact of Calibration of pipette on Quality Control Results


Null

Author Details : Ketan K Mangukiya, Mittal Panchal

Volume : 3, Issue : , Year : 2016

Article Page : 28-30


Suggest article by email

Get Permission

Abstract

Background: Majority of biochemical investigations requires pipette for delivery of reagents and samples in clinical laboratory so that pipette that is used for the same must be properly calibrated for accurate results.
Aims: The aim of this study to investigate that the impact of automatic pipette calibration on quality control results.
Methodology: Total 6 pipettes are used in our study. Among them 3 pipettes are with volume 1.0 ml and another 3 are with volume 0.1 ml, all pipettes were examined for work calibration. Distilled water weighing method is used for calculation of accuracy and precision of pipettes. Control sera of glucose and uric acid is used for determining accuracy and Precision of pipettes.
Result & Conclusion: Uncalibrated pipette had high effect on accuracy of control data as well as patients results therefore it is recommended that all pipette that is used in clinical laboratory must be periodically checked for calibration.

Keywords:
Accuracy, Calibration, Gujarat, Pipette, Precision


How to cite : Mangukiya K K, Panchal M, Impact of Calibration of pipette on Quality Control Results. Int J Clin Biochem Res 2016;3():28-30

This is an Open Access (OA) journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.