International Journal of Clinical Biochemistry and Research

Print ISSN: 2394-6369

Online ISSN: 2394-6377

CODEN : IJCBK6

International Journal of Clinical Biochemistry and Research (IJCBR) open access, peer-reviewed quarterly journal publishing since 2014 and is published under auspices of the Innovative Education and Scientific Research Foundation (IESRF), aim to uplift researchers, scholars, academicians, and professionals in all academic and scientific disciplines. IESRF is dedicated to the transfer of technology and research by publishing scientific journals, research content, providing professional’s membership, and conducting conferences, seminars, and award more...

Article type

Original Article


Article page

376- 379


Authors Details

Dipti Basavaraj Sanakal, Sagar Chandrakant Mhetre, Ramesh S Patil, Abdul Kayyum Shaikh


View Article As

 


Downlaod Files

   






Article statistics

Viewed: 1824

PDF Downloaded: 888


Reliability of blood gas analyzer for the measurement of electrolytes - A comparative study


Original Article

Author Details : Dipti Basavaraj Sanakal, Sagar Chandrakant Mhetre, Ramesh S Patil, Abdul Kayyum Shaikh

Volume : 3, Issue : 4, Year : 2016

Article Page : 376-379


Suggest article by email

Get Permission

Abstract

Background: Electrolyte and blood gas analysis are essential for most patients who are admitted in the casualty. Routinely simultaneous arterial and venous sample would be sent to the laboratory for blood gas analysis and serum electrolytes respectively. Electrolytes can also be measured   by the blood gas analyzer. We wanted to find out if these values were reliable.
Methods: This study has been checked and permitted by the hospital ethics committee. We obtained two samples, one venous and one arterial sample from patients in the intensive care unit (ICU) and casualty of Ashwini rural medical college, hospital and research centre, (ARMCH&RC) Kumbhari Solapur.
The arterial blood samples were analyzed on the ABOTT arterial blood gas analyzer (ABG) and venous samples were analyzed on PROLYTE electrolyte autoanalyzer (AA). Both instruments were placed in our biochemistry laboratory.
Results: A total of 200 samples were analyzed. The mean ABG sodium value was 132.0 mmol/L (SD 9.82) and the mean AA sodium (Na+) was 135.7 mmol/L (SD 9.99). The mean AA sodium value was more than mean ABG sodium but statistically significant difference was not found (p=0.18).
The mean ABG potassium (K+) value was 3.4 mmol/L (SD 0.68) and the mean AA potassium (K+) was 3.8 mmol/L (SD 0.79). The mean AA potassium value was more than mean ABG potassium and found statistically significant difference (p=0.02).
Conclusion: We found that there was no significant difference in the sodium values measured by the ABG analyzer and the auto-analyzer. However, there was a significant difference in the potassium values measured by these instruments.

Keywords: Electrolytes; Arterial blood gas analyzer (ABG); Autoanalyzer (AA)


How to cite : Sanakal D B, Mhetre S C, Patil R S, Shaikh A K, Reliability of blood gas analyzer for the measurement of electrolytes - A comparative study. Int J Clin Biochem Res 2016;3(4):376-379

This is an Open Access (OA) journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.