International Journal of Clinical Biochemistry and Research

Print ISSN: 2394-6369

Online ISSN: 2394-6377

CODEN : IJCBK6

International Journal of Clinical Biochemistry and Research (IJCBR) open access, peer-reviewed quarterly journal publishing since 2014 and is published under auspices of the Innovative Education and Scientific Research Foundation (IESRF), aim to uplift researchers, scholars, academicians, and professionals in all academic and scientific disciplines. IESRF is dedicated to the transfer of technology and research by publishing scientific journals, research content, providing professional’s membership, and conducting conferences, seminars, and award more...

Article type

Original Article


Article page

220- 224


Authors Details

Ayaz Khurram Mallick, Marya Ahsan


View Article As

 


Downlaod Files

   






Article statistics

Viewed: 1466

PDF Downloaded: 1016


A comparative study of glucose concentration determined from venous plasma sample and capillary blood sample


Original Article

Author Details : Ayaz Khurram Mallick, Marya Ahsan

Volume : 4, Issue : 3, Year : 2017

Article Page : 220-224


Suggest article by email

Get Permission

Abstract

Context: Self-monitoring of glucose levels with a hand held glucometer is very popular among the diabetic patients and doctors as it is less traumatic and gives instant results. Capillary blood samples are commonly used for self-monitoring in comparison to venous plasma sample which are used in the diagnostic laboratories.
Aims: This study was done to compare the glucose levels of a capillary blood sample determined by a glucometer with the glucose level determined by Glucose Oxidase Peroxidase (GOD-POD) method using a venous plasma sample.
Setting and Design: This experimental study was carried out at Rohilkhand Medical College and Hospital.
Methods and Material: Two hundred patients were randomly selected for the study. Their venous blood samples were collected in a fluoride containing blood collection tube. The plasma obtained was subjected to GOD-POD test. Simultaneously their capillary blood glucose was also determined by a glucometer.
Statistical analysis used: The data collected was analyzed using SPSS 17.0 version. Data was expressed as mean ± SD. Significance of mean was analyzed using paired t test. P value < 0.05 was considered significant.
Result: Glucose levels determined by a glucometer and GOD-POD method showed very good correlation (r=0.958; p<0.001). However, at very high and low glucose levels, the glucometer significantly under estimated (p=0.014) and over-estimated (p<0.001) the glucose levels respectively.
Conclusion: Although glucometers are effective for routine monitoring of glucose levels, they should be used cautiously in emergency conditions.

Keywords: Glucose levels, GOD-POD method, Glucometer


How to cite : Mallick A K, Ahsan M, A comparative study of glucose concentration determined from venous plasma sample and capillary blood sample. Int J Clin Biochem Res 2017;4(3):220-224

This is an Open Access (OA) journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.