International Journal of Clinical Biochemistry and Research

Print ISSN: 2394-6369

Online ISSN: 2394-6377

CODEN : IJCBK6

International Journal of Clinical Biochemistry and Research (IJCBR) open access, peer-reviewed quarterly journal publishing since 2014 and is published under auspices of the Innovative Education and Scientific Research Foundation (IESRF), aim to uplift researchers, scholars, academicians, and professionals in all academic and scientific disciplines. IESRF is dedicated to the transfer of technology and research by publishing scientific journals, research content, providing professional’s membership, and conducting conferences, seminars, and award more...

Article type

Original Article


Article page

334- 337


Authors Details

Shalini L., Santhi Silambanan


Article Metrics


View Article As

 


Downlaod Files

   






Article statistics

Viewed: 1244

PDF Downloaded: 727


A study on the assessment of stability of glucose concentrations in serum separator gel tubes


Original Article

Author Details : Shalini L., Santhi Silambanan

Volume : 5, Issue : 2, Year : 2018

Article Page : 334-337

https://doi.org/10.18231/2394-6377.2018.0069



Suggest article by email

Get Permission

Abstract

Introduction: Increasing diabetic burden in India has shifted our health care focus towards its prevention and management of diabetes mellitus. Often multiple investigations are requested by the physicians to assess diabetic status. Conventionally fluoride tubes (glycolytic inhibitor) have been used for glucose analysis in clinical laboratories. Preservatives used here make it unsuitable to measure other key clinical chemistry analytes, for which serum separator gel tubes(SST) are being employed. On centrifugation of SST tubes, a thick gel barrier is formed separating the serum from the cells. This gel barrier prevents utilization of glucose by the cells, hence researchers have considered using it for glucose measurement. This study has been done to assess the integrity of serum samples obtained from SST tubes for analysis of glucose.
Materials and Methods: A single centre cross-sectional study was done, with 140 paired samples collected in fluoride and SST tubes from patients attending Sri Ramachandra Laboratory Services. Samples were analyzed by glucose oxidase-peroxidase method in Siemens ADVIA 1800 Clinical Chemistry at 1 hr, 6 hrs and 24 hrs of sample collection.
Results:Statistical analysis showed no significant change in glucose concentrations measured in fluoride against serum samples. Paired t-test results showed no statistical significance between the tubes, irrespective of time. Bland Altman plot analysis revealed that serum values were not significantly different from fluoride samples throughout the study period. Intraclass correlation (ICC) showed correlation of 0.999 between serum and plasma at 1 hr, 6 hrs and 24 hrs from time of collection.
Conclusions: Serum from SST tubes can be used to measure blood glucose instead of standard accepted fluoride plasma samples.

Keywords: Diabetes mellitus, Fluoride tube, Serum separator gel tube, Analytes, Glucose, Gel barrier


How to cite : Shalini L., Silambanan S, A study on the assessment of stability of glucose concentrations in serum separator gel tubes. Int J Clin Biochem Res 2018;5(2):334-337

This is an Open Access (OA) journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.