Validation and performance evaluation of biochemical assays in pleural fluid, ascitic fluid, and cerebrospinal fluid: A comprehensive precision, sensitivity, accuracy, and linearity study
- Author Details:
-
Jayesh Prabhakar Warade *
-
Ashish Anjankar
Background: The validation of biochemical assays in pleural fluid, ascitic fluid, and CSF is essential for accurate diagnosis, disease monitoring, and treatment evaluation. This study aims to assess the precision, sensitivity, accuracy, and linearity of biochemical assays performed on these fluids using various analytes.
Objectives: The primary objective of this study is to evaluate the performance characteristics of biochemical assays in pleural fluid, ascitic fluid, and CSF, specifically focusing on precision (CV%), sensitivity (CV% at the lower detection limit), accuracy (recovery rates), and linearity (correlation coefficients).
Materials and Methods: This in-house validation study utilized a range of analytes, including glucose, cholesterol, triglyceride, albumin, urea, creatinine, total protein, amylase, LDH, ADA, and lipase. The precision of each assay was measured by calculating the CV% at both low and high concentrations. Sensitivity was assessed by evaluating the lowest concentration detectable. Accuracy was determined by calculating recovery rates, while linearity was assessed through correlation coefficients.
Results: The precision study showed that the CV% for all analytes across the three body fluids was consistently below 5%, with pleural fluid and ascitic fluid demonstrating excellent precision. Sensitivity studies confirmed reliable detection limits for the assays, with CV% below 15% for most analytes. The accuracy study revealed that recovery rates for analytes ranged from 93.5% to 108.6%, indicating strong alignment with expected values. The linearity study demonstrated high correlation coefficients between 0.95 and 1.0 for most analytes across pleural and ascitic fluids, ensuring reliable results even at higher concentrations.
Conclusion: This study successfully validates the use of biochemical assays for pleural fluid, ascitic fluid, and CSF. The results indicate that these assays provide reliable and consistent diagnostic information, making them valuable tools for clinicians. The findings support the use of these assays in clinical settings for disease monitoring and treatment response evaluation.
References
- Janssens PMW. Recognizing and differentiating uncommon body fluids: Considerations and tools for a proper practical approach. Clin Chim Acta. 2017;471:6-11. https://doi:10.1016/j.cca.2017.05.005
[Google Scholar] - Chubb SP, Williams RA. Biochemical analysis of pleural fluid and ascites. Clin Biochem Rev. 2018;39(2):39-50.
- Hrishi AP, Sethuraman M. Cerebrospinal fluid (CSF) analysis and interpretation in neurocritical care for acute neurological conditions. Indian J Crit Care Med. 2019;23(2):S115-9. https://doi:10.5005/jp-
[Google Scholar] journals-10071-23187. - Medlab Magazine. Validating the performance of body fluid specimens. Medlabmag.com. 2016; Available from: https://www.medlabmag.com/article/1291
- Clinical and Laboratory Standards Institute. Analysis of body fluids in clinical chemistry; approved guideline. CLSI document C49-A. Wayne (PA): CLSI; 2007.
- Block DR, Ouverson LJ, Wittwer CA, Saenger AK, Baumann NA. An approach to analytical validation and testing of body fluid assays for the automated clinical laboratory. Clin Biochem. 2018;58:44-52. https://doi:10.1016/j.clinbiochem.2018.05.002
[Google Scholar] - Block DR, Algeciras-Schimnich A. Body fluid analysis: Clinical utility and applicability of published studies to guide interpretation of today’s laboratory testing in serous fluids. Crit Rev Clin Lab Sci. 2013;50(4-5):107-24. https://doi:10.3109/10408363.2013.844679
[Google Scholar] - Eigsti RL, Krasowski MD, Vidholia A, Merrill AE. Data on interference indices in body fluid specimens submitted for clinical laboratory analysis. Data Brief. 2020;30:105408. https://doi:10.1016/j.dib.2020.105408
[Google Scholar] - Oris C, Landrieaux L, Brailova M, Roszyk L, Rozand I, Sapin V, et al. Validation de méthodes des dosages de 13 paramètres biochimiques (ACE, CA 19-9, amylase, lipase, sodium, potassium, chlorure, créatinine, glucose, protéines, albumine, LDH, triglycérides) dans un panel de liquides biologiques. Ann Biol Clin (Paris). 2020;78(1):93-107. https://doi:10.1684/abc.2020.1524
[Google Scholar] - Cho H, Jung J, Lee JJ, Oh EJ, Kim Y, Chae H. Body fluid matrix effect evaluation on the Hitachi Labospect 008 system. Clin Biochem. 2021;95:73-6. https://doi:10.1016/j.clinbiochem.2021.05.005
[Google Scholar] - Lo SY, Saifee NH, Mason BO, Greene DN. Filling in the gaps with non-standard body fluids. Pract Lab Med. 2016;5:24-31. https://doi:10.1016/j.plabm.2016.03.003
[Google Scholar] - Raviraj, Henry RA, Rao GG. Determination and validation of a lower cut off value of cerebrospinal fluid adenosine deaminase (CSF-ADA) activity in diagnosis of tuberculous meningitis. J Clin Diagn Res. 2017;11(4):22-4. https://doi:10.7860/JCDR/2017/25823.9625
[Google Scholar] - Arrigo C, Aloisio E, Rovegno L, Dolci A, Panteghini M. The laboratory investigation of pleural fluids: An update based on the available evidence. Ann Clin Biochem. 2023;60(4):228-35. https://doi:10.1177/00045632231157547
[Google Scholar] - Rodríguez Vargas BO, Monge Salgado E, Montes Teves P. Salazar Ventura S, Guzmán Calderón E. Eficacia de la gradiente de albúmina sangre-ascític y los análisis de proteínas en líquido ascítico en el diagnóstico de ascitis hipertensiva portal. Rev Gastroenterol Peru. 2014;34(1):23-8.
- N Boyert, A J McShane, A-097 Body Fluid Validation for 11 Analytes on a Chemistry Analyzer, Clinical Chemistry, Volume 70, Issue Supplement_1, 2024, hvae106.096, https://doi.org/10.1093/clinchem/hvae106.096
[Google Scholar]
How to Cite This Article
Vancouver
Warade JP, Anjankar A. Validation and performance evaluation of biochemical assays in pleural fluid, ascitic fluid, and cerebrospinal fluid: A comprehensive precision, sensitivity, accuracy, and linearity study [Internet]. Int J Clin Biochem Res. 2025 [cited 2025 Oct 26];12(3):149-155. Available from: https://doi.org/10.18231/j.ijcbr.13447.1759901703
APA
Warade, J. P., Anjankar, A. (2025). Validation and performance evaluation of biochemical assays in pleural fluid, ascitic fluid, and cerebrospinal fluid: A comprehensive precision, sensitivity, accuracy, and linearity study. Int J Clin Biochem Res, 12(3), 149-155. https://doi.org/10.18231/j.ijcbr.13447.1759901703
MLA
Warade, Jayesh Prabhakar, Anjankar, Ashish. "Validation and performance evaluation of biochemical assays in pleural fluid, ascitic fluid, and cerebrospinal fluid: A comprehensive precision, sensitivity, accuracy, and linearity study." Int J Clin Biochem Res, vol. 12, no. 3, 2025, pp. 149-155. https://doi.org/10.18231/j.ijcbr.13447.1759901703
Chicago
Warade, J. P., Anjankar, A.. "Validation and performance evaluation of biochemical assays in pleural fluid, ascitic fluid, and cerebrospinal fluid: A comprehensive precision, sensitivity, accuracy, and linearity study." Int J Clin Biochem Res 12, no. 3 (2025): 149-155. https://doi.org/10.18231/j.ijcbr.13447.1759901703