Impact of Calibration of pipette on Quality Control Results
- Author Details:
-
Ketan K Mangukiya
-
Mittal Panchal
Background: Majority of biochemical investigations requires pipette for delivery of reagents and samples in clinical laboratory so that pipette that is used for the same must be properly calibrated for accurate results.
Aims: The aim of this study to investigate that the impact of automatic pipette calibration on quality control results.
Methodology: Total 6 pipettes are used in our study. Among them 3 pipettes are with volume 1.0 ml and another 3 are with volume 0.1 ml, all pipettes were examined for work calibration. Distilled water weighing method is used for calculation of accuracy and precision of pipettes. Control sera of glucose and uric acid is used for determining accuracy and Precision of pipettes.
Result & Conclusion: Uncalibrated pipette had high effect on accuracy of control data as well as patients results therefore it is recommended that all pipette that is used in clinical laboratory must be periodically checked for calibration.
Keywords: Accuracy, Calibration, Gujarat, Pipette, Precision
References
- Standard operating procedure Air displacement pipette Calibri, Denver: university of Colorado Shared analytical services laboratory, 2003.
- http://www.radwag.com/pliki/artykuly/gravimetric_deter mination_of_pipette_errors.pdf.
- Calibration manual for Finn Pipettes: http://www.pipette- calibration.net/pipette-calibrationfiles/Guide-To- Pipetting-2.pdf.
- Y.C. Chang. Laboratory manual of Clinical Biochem. Res. Lab Dept. of Res. & Edu. Apr. 14, 2007.
- Automatic micro pipettes, available at: Slc.umd.umich. edu.slconline micro pipette,2015.
- Density of water in 20°C based on recent experimental reports. Metrologia, 2001;38:301-309.
How to Cite This Article
Vancouver
Mangukiya KK, Panchal M. Impact of Calibration of pipette on Quality Control Results [Internet]. Int J Clin Biochem Res. 2016 [cited 2025 Oct 12];3(1):28-30. Available from: https://doi.org/
APA
Mangukiya, K. K., Panchal, M. (2016). Impact of Calibration of pipette on Quality Control Results. Int J Clin Biochem Res, 3(1), 28-30. https://doi.org/
MLA
Mangukiya, Ketan K, Panchal, Mittal. "Impact of Calibration of pipette on Quality Control Results." Int J Clin Biochem Res, vol. 3, no. 1, 2016, pp. 28-30. https://doi.org/
Chicago
Mangukiya, K. K., Panchal, M.. "Impact of Calibration of pipette on Quality Control Results." Int J Clin Biochem Res 3, no. 1 (2016): 28-30. https://doi.org/