International Journal of Clinical Biochemistry and Research (IJCBR) open access, peer-reviewed quarterly journal publishing since 2014 and is published under auspices of the Innovative Education and Scientific Research Foundation (IESRF), aim to uplift researchers, scholars, academicians, and professionals in all academic and scientific disciplines. IESRF is dedicated to the transfer of technology and research by publishing scientific journals, research content, providing professional’s membership, and conducting conferences, seminars, and award more...

Article type

Original Article


Article page

356- 362


Authors Details

K Maria Kumar*, K Vijayakumari, K Sri Lakshmi


Article Metrics


View Article As

 


Downlaod Files

   






Article statistics

Viewed: 1386

PDF Downloaded: 1369


A prospective comparative study of glucose estimation by hexokinase and glucose oxidase-peroxidase methods


Original Article

Author Details : K Maria Kumar*, K Vijayakumari, K Sri Lakshmi

Volume : 6, Issue : 3, Year : 2019

Article Page : 356-362

https://doi.org/10.18231/j.ijcbr.2019.078



Suggest article by email

Get Permission

Abstract

Aims and Objectives: Error analysis of the comparative study of glucose estimation with the comparison
and regression analysis.
Introduction: Glucose is the major carbohydrate found in the blood and a chief source of energy in human
body. Estimation of glucose is very important in clinical diagnosis of hyperglycemia, hypoglycemia and
normoglycemia. Accuracy of test results are very important, requires considerable effort and cost. For
introducing in the laboratory the test method has to be evaluated by the process called quality control.
Method selection and evaluation is important in improving efficiency.
Materials and Methods: The preferred sample for Glucose assay is fasting serum or plasma. A total of
105 samples are analysed for the quantitative estimation of blood glucose by the reference method.
Discussion and Results: In the present study comparison between a test method (Glucose oxidaseperoxidase)
and a reference method (Hexokinase), the errors are less than TEa. Mean value of Hexokinase
method is 311.36 137.315 and mean value Glucose oxidase-peroxidase method is 291.87 133.412.
p-value on comparison = 0.00014. With this the test method can be an alternative to reference method in
terms of accuracy and precision and analytical sensitivity and specificity.
Conclusion: Glucose oxidase-peroxidase method is also a suitable method for the quantitative estimation
of glucose in need of clinical requirements.

Keywords: Serum glucose, Comparison and correlation, Glucose oxidase -peroxidase, Hexokinase, Method evaluation, Quality control.


How to cite : Kumar K M, Vijayakumari K, Lakshmi K S, A prospective comparative study of glucose estimation by hexokinase and glucose oxidase-peroxidase methods. Int J Clin Biochem Res 2019;6(3):356-362

This is an Open Access (OA) journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.